- The sensor was previously exposed to very wet air and needs recovery time.
- Contamination such as oil, particles, or chemicals is slowing moisture exchange at the sensing surface.
- The sample flow is too low or stagnant.
- The sensor is installed in a dead-leg or in a point with poor air exchange.
- The sensing element has aged and no longer meets its original response specification.
- Verify that sufficient sample flow reaches the sensor and that the measuring point is representative.
- Check whether the sensor has been exposed to condensate or oil contamination.
- Compare response after a process step change with the expected sensor behavior from the datasheet.
- Clean or service the sampling system if applicable, but do not damage the sensing element.
- Recalibrate or replace the sensor if long-term drift or aging is suspected.